Brewer Access High-Low User Manual
Page 19

Document # 100452 RevL
19
Printed in USA © 2013
Step #
1
2
3
4
5
6
7
* No relay clicking in control box. Table then functions after about 30 minutes.
Front cabinet switches; 1 or more
of 6 is not closing when pressed.
Press each one and listen for click
when pressed and again when
released. Replace as necessary.
Start Here
Complete the operational test
Test fails
Test passes
Test Complete
No movement
Some
movement
Go to step 2.1
Go to step 2.2
Does the drawer warmer heat
properly? Does outlet work?
Yes or N/A
No
Call 1 888-Brewer1 for technical
assistance.
Call 1 888-Brewer1 for technical
assistance.
Fixed
Still bad
Seat back switches; 1 or more of 4
is not opening when pressed.
Press each one and listen for click
when pressed and again when
released. Replace as necessary.
Fixed
Yes
No
Return to step 1
Control box is over temperature.
Allow 30 minute cool down.
Control box resets automatically.
Yes
No
No
Yes
Is there excessive noise or other
problems which are not yet
identified?
Go to step 5.1 for Heater
Go to step 6.1 for Outlet
Go to step 7.1
Table stops after period of heavy
use. *see below
Still bad
Do all powered functions work?
Yes
No
Do the 6 front limit switches
work properly?
Yes
No
Do the 4 seat back limit switches
work properly?
6000 SERIES TABLE TROUBLESHOOTING CHART