Warning, Caution – Brewer Access High-Low User Manual
Page 4

Document # 100452 RevL
4
Printed in USA © 2013
WARNING
When performing a cauterization or similar treat-
ment, the patient must be insulated from the metal
portions of the table by non-conductive material.
Failure to do so may result in electrical shock or
burns to the patient.
WARNING
Use 115 VAC, 60 Hz as noted on the rating label.
Failure to do so may result in electrical shock to
personnel and will result in damage to the table.
WARNING
Do not use this table in an explosive or oxygen-
enriched atmosphere. Failure to do so may result
in serious personal injury or death.
CAUTION
Do not use any power supply other than that
listed on the rating label (Figure 1). Failure to do
so may result in serious injury and/or equipment
damage.
SPECIFICATIONS
Fuse Rating:
Control Box Fuse Rating, 120V Models ............ 250V/2.5A
Option/Warmer Fuse Rating .............................. 250V/1.0A
Option/Warmer & Outlet Fuse Rating ................ 250V/3.0A
Duty Cycle (10%)
.................... 2 minutes on / 18 minutes off (motor run time)
Power Cord............... Extends 70 in. (Minimum) from table.
120V Models: 18 AWG / 3 conductor, SJT grey jacketed
junior hard service with hospital plug or detachable
power cord.
Type of protection
against electrical shock
Model 6000, 6500 and 6800 ........ Class 2 Double Insulated
Models 6001, 6501 and 6801 ................. Class 1 Grounded
Type of protection
against electrical shock ............................... B applied parts
Type of protection against ingress of water……. Ordinary
Can accept paper rolls of
18 in. x 3.0 in.
.................................................................. or 21 in. x 3.0 in.
Certifications* ..............................UL60601-1; IEC60601-1;
..................................................CAN-CSA C22.2 No. 601.1
Operating Conditions:
Temperature Range ...........................................65° to 85°F
Relative Humidity..............................................10% to 90%
Transportation and Storage
Temperature .................................................-20°F to 150°F
Humidity............................................................10% to 90%
* UL classified in the United States and Canada
per the standards listed above.
Weight of Table (
Models 6000, 6001, 6500 & 6501
) .. 395 lbs.
Weight of Table (
Model 6800
) ..................................
423 lbs
Weight of Table (
Model 6801
) ..................................
432 lbs
Load Rating (maximum):
- Seat/Back 6800/6801 .......................................... 700 lbs.
- Seat/Back 6000/6001/6500/6501 ........................ 450 lbs.
- Leg Support .......................................................... 50 lbs.
- Leg Extension ....................................................... 50 lbs.
- Pass Through Work Surface ................................. 15 lbs.
- Grab Bars ............................................................ 100 lbs.
Back Section Range .................0° (horizontal) to +80° ± 5°
Table Top Height Range .............18 ± 0.5 in. to 37 ± 1.0 in.
Pelvic Tilt Range
(Models 6001/6501/6801 Only) ....0° (horizontal) to 5° (up)
Upholstery Dimensions:
Upholstered Top ........................ 28.0 in. wide x 60 in. long
Upholstered Leg Support……15.25 in. long x 18.5 in. wide
Upholstered Leg Extension…....6.0 in. long x 15.5 in. wide
Overall Length ................................................... 60 in. long
With Leg Support Extended............................... 72 in. long
With Leg Extension Extended ........................... 78 in. long
Electrical Requirement
6000/6500/6800
........
115 VAC nominal, 60 HZ, 5.8 amps max.
6001
.........................
115 VAC nominal, 60 HZ, 6.3 amps max.
6001/6501/6801 W/Outlet
...
115 VAC nominal, 60 HZ, 9.3 amps
max.
ELECTRICAL REQUIREMENTS
CAUTION
This product has been evaluated with respect to
electrical shock, fire, and mechanical hazards only
in accordance with UL60601-1; IEC60601-1;
CAN/CSA C22.2 No. 601.1
Electromagnetic Interference
This product is designed and built to minimize electromag-
netic interference with other devices; however, if interfer-
ence is noticed between another device and this product,
remove the interfering device from the room or plug this
product into an isolated circuit.