OPERATION (continued)
CAUTION
Do not use any part of the stirrup for attachment of any
device (Figure 26). Warranty will be voided. Use support
adapters to attach clamps, then attach knee crutches to
clamps. If adapters are not currently installed, order PN
100550 (See page 24 for accessories).
Figure 24. Stirrups
Figure 25. Stirrup Locking Mechanism
Figure 26. Affix No Attachments!
Storing the Stirrups
1. To store the stirrups in the table, grasp end of stirrup
and pull straight out to its full extension.
2. Fold stirrup down against bar, then rotate it to the
inner most lateral position and slide it back into the
stowed position.
Close Up of Locking Mechanism
WARNING
Failure to engage the lateral locking mechanism
could allow patient to lose balance resulting in
personal injury to patient.
Adjustable Stirrups
Operating the Stirrups (Figure 24)
1. To operate the stirrups, grasp end of stirrup and pull
straight out of the table. Lift up slightly while sliding
the stirrup out.
NOTE: Do not apply any downward pressure when pull-
ing stirrup from table or it may be difficult to move. This
is normal and is part of the stirrup friction locking mecha-
nism
2. Unfold stirrup upward to the fully open position.
3. Adjust the stirrup to the desired length by sliding it in
or out.
Stirrup Lateral Adjustment
1. Lift end of stirrup and then rotate outward to
desired position.
2. When the desired position is achieved, lower the stir-
rup to engage the lateral locking mechanism (Figure
25).
3. Check that lateral locking mechanism is engaged by
attempting to rotate stirrup assembly without lifting on
the stirrup end.
See enlargement
(Figure 20)