2.1
a
b
c
d
e
f
g
Return to step 1
Step 2.1 continue here
Call 1 888-Brewer1 for technical
assistance.
Front cabinet switch (one of 6) is
stuck on. (closed) Unplug table
power. Isolate faulty switch or wire
using a continuity tester. Replace
or repair.
Is green light on at back side of
control box?
No
Yes
Is the table stuck in the up
position?
No
Still bad
Fixed
No heater or No
Table does not go down. Control
box relay clicks in down direction.
When front bottom switch plate is
lifted there is no relay click.
No
Yes
Call 1 888-Brewer1 for technical
assistance.
Return to step 1
Return to step 1
Check for loose wire connections
at terminal block.
Return to step 1
Return to step 1
If table has a heater, is the switch
illuminated when turned on?
Yes
Still bad
Fixed
Press foot control pedal. Does
relay in control box click?
Yes
Replace 2 control box fuses.
Still bad
Is on/off switch on back of cabinet
turned on?
Yes
No
Check for loose plug in wall outlet
or back of table. Check for power
at outlet.
Yes
No
Fixed
Replace control box.
Still bad
Fixed
Ok
Turn on switch, then return to step 1
Fixed
6000 SERIES TABLE TROUBLESHOOTING CHART (continued)