beautypg.com

Safety, Regulatory information – Canon CR-2 User Manual

Page 7

background image

7

2

Safety

Regulatory Information

The following sections list the classifications applicable to the CR-2 and the directives and standards

that the CR-2 complies with.

Device Classification

Protection against electric shock

Class I equipment

Degree of protection against electric shock

Type B applied parts

Degree of protection against ingress of water

IPX0

Degree of safety of application in the presence of a

flammable anaesthetic mixture with air or with oxygen or

nitrous oxide

Not suitable

Mode of operation

Continuous operation

Directives and Standards

USA and Canada

UL 60601-1

CAN/CSA C22.2 No.601.1

IEC 60601-1: 1988+A1: 1991+A2: 1995

IEC 60601-1: 2005

Medical electrical equipment – Part 1: General

requirements for safety

IEC 60601-1-1: 2000

Medical electrical equipment – Part 1-1: General

requirements for safety – Collateral standard: Safety

requirements for medical electrical systems

IEC60601-1-2: 2001+A1: 2004

Medical electrical equipment – Part 1-2: Collateral

standard: Electromagnetic compatibility – Requirements

and tests

IEC 60601-1-4: 1996+A1: 1999

Medical electrical equipment – Part 1-4: Collateral

standard: Programmable electrical medical systems

ISO 15004-1: 2006

Ophthalmic instruments – Part 1: General requirements

applicable to all ophthalmic instruments

ISO 15004-2: 2007

Ophthalmic instruments – Part 2: Light hazard protection

ISO 10993-1: 2009

ISO 10993-5: 2009

ISO 10993-10: 2009

Biological evaluation of medical devices – Evaluation and

testing