beautypg.com

Troubleshooting – Canon CR-2 User Manual

Page 40

background image

40

If any of the symptoms listed here occurs or if a warning appears in the Rics NM 2 while using the

CR-2, try the relevant remedy described below.

If these remedies do not eliminate the symptoms or the warning, turn off the power and contact

your sales representative or local Canon dealer. Be ready to provide a detailed description of the

symptoms and message.

Symptom

Cause and remedy

The power lamp does not light

when the power is turned on.

The power cord has been disconnected. Turn off the power to

the retinal camera and reconnect the power cord.

The power cord has been cut. Contact your sales

representative or local Canon dealer.

5 short beeps sound and the

operation panel lamps blink when

the power is turned on.

A system error has occurred. Contact your sales representative

or local Canon dealer.

The observation monitor is dim

and hard to see.

Adjust the brightness of the observation monitor. See the Rics

NM 2 manual.

The observation monitor does not

turn on.

Turn on the digital camera.

An error appears on the

observation monitor.

A malfunction has been detected in the digital camera. Turn off

the power to the retinal camera and then turn it on again.

Err 06

Sensor cleaning is not possible.

Err 70

Photography is not possible due to an error.

The working distance dots are

hard to see.

The observation light is too bright. Reduce the observation light

brightness.

The split lines are hard to see.

The working distance dots are not

visible.

The retinal camera and the eye are not aligned. Align them

correctly.

Flare appears in the image.

The retinal camera and the eye are not aligned. Align them

correctly. If flare still appears, refer to items 1 through 4 below

and try taking the photo again.

1)

White flare appears around

the edges of the image.

The retinal camera and the patient’s eye are separated. Move

the objective lens closer to the patient’s eye.

2)

Orange flare appears around

the edges of the image.

The retinal camera and the patient’s eye are too close. Move

the objective lens away from the patient’s eye.

3)

White flare appears on the

left side of the image.

The retinal camera is too close to the left side of the patient’s

eye. Move the objective lens to the right.

4)

White flare appears on the

right side of the image.

The retinal camera is too close to the right side of the patient’s

eye. Move the objective lens to the left.

The split lines are not visible.

The patient's pupil is not dilated enough. Dilate the patient’s

pupil. If the pupil is still too small, enable the small pupil

photography function.

Eyebrows or eyelashes are blocking the pupil. Instruct the

patient to open his/her eye wider.

The patient's eye is cloudy. Focusing cannot be performed with

the split lines.

The diopter compensation knob is set to either negative or

positive. Set it to the 0 position.

The split line lamps are not lit.

7

Troubleshooting