7 Troubleshooting

41

Symptom

Cause and remedy

The split lines do not align.

The diopter patient's eye is outside the range of –10 to +15 D.

Use the diopter compensation function, turn the focus ring, and

focus on the position where the images of the retina look clear

at the same time.

Cannot capture good images.

The patient's pupil is not dilated enough. Dilate the patient’s

pupil. If the pupil is still too small, enable the small pupil

photography function.

The retinal camera and the eye are not aligned. Align them

correctly.

Focus the image.

Eyebrows or eyelashes are blocking the pupil. Instruct the

patient to open his/her eye wider.

The patient's eye is cloudy. Clear images cannot be captured if

this is the case.

Adjust the flash intensity, patient's diopter, and small pupil

photography function.

The objective lens is dirty. Clean the objective lens (see page 35).

The patient is wearing contact lenses. Instruct the patient to

remove his/her contact lenses.

Black spots or any artifacts

appear in the same place in every

captured image.

There is dirt or dust on the objective lens. Clean the objective

lens (see page 35).

Dirt or dust is stuck to the imaging sensor of the digital camera.

Use the self-cleaning sensor unit to clean it. (see page 36).

White spots are visible on the

captured image.

The objective lens is dirty. Clean the objective lens (see page 35).

The patient's eyelashes are interfering with the image. Instruct

the patient to open his/her eye wider, and capture another

image.