Canon CR-2 User Manual
Page 11
2 Safety
11
CAUTION
Prohibited
Do not install in locations exposed to water, steam, moisture
or dust.
Doing so may cause problems or malfunctions.
Prohibited
Do not install in locations exposed to salt, sulfur or corrosive
gas.
Doing so may result in corrosion of the instrument, problems or malfunctions.
Prohibited
Do not install in locations that are unstable or exposed to
vibration.
The vibration may knock over the instrument or the instrument may become
unbalanced and fall, resulting in a malfunction or injury.
Prohibited
Do not place anything near the power plug.
To make it easy to unplug the power plug at any time, avoid putting any obstructions
near the AC outlet.
Fire or electric shock may result if the power plug is not unplugged during an
emergency.
Prohibited
Do not block the vent holes.
Doing so may cause the internal temperature to rise and result in fire.
Prohibited
Do not place your hands or fingers on the base.
Your hand or fingers may be pinched and injured when the stage moves to either
side.
Similarly, instruct the patient not to place his/her hands or fingers on the base.
Prohibited
Do not hold the head rest or digital camera when moving the
CR-2.
When carrying the CR-2, be sure to tighten the stage lock, hold the carrying
indentations at the front and rear of the base panel, and keep the CR-2 level. Do not
hold it by the digital camera or the head rest or any other parts, as they may come
off and result in injury.
Hold the CR-2 when connecting or disconnecting a cable.
For safety reasons, when connecting or disconnecting the power cord or any cable,
be sure to hold the main unit. Otherwise, the main unit may fall over, possibly
causing injury.
Ensure that the entire system conforms to IEC 60601-1-1.
Use the computer, monitor, and other equipment that conform to the system
standard IEC 60601-1 or IEC 60950-1 for the Digital Retinal Camera CR-2. Be sure
that the entire system conforms to IEC 60601-1-1. Be sure to also use an isolation
transformer conforming to IEC 60601-1 when a computer or monitor conforming to
IEC 60950-1 is used.
Otherwise, electric shock may occur. For details, please contact your sales
representative or local Canon dealer.