Appendix

43

EMC (Electromagnetic Compatibility)

The CR-2 is designed and tested to comply with IEC 60601-1-2 (EN 60601-1-2), the applicable regulations regarding EMC

for medical devices. The CR-2 needs to be installed and put into service according to the EMC information stated as

follows.

If this equipment causes harmful interference to other devices, which can be determined by turning the equipment off

and on, the user is encouraged to try to correct the interference by one or more of the following measures:

• Reorient or relocate the receiving device.

• Increase the separation between the equipment.

• Connect this device into an outlet on a circuit different from that to which the other devices are connected.

If the problem cannot be solved with the above measures, stop using this equipment and consult your sales representative

or local Canon dealer.

Precautions on EMC

1.

Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service

according to the EMC information provided in the manual.

2.

Portable and mobile RF communications equipment can affect medical electrical equipment.

3. Information regarding the cable affecting EMC is as follows.

To maintain the optimum EMC performance, use only the designated cables.

Name

Type

Length

Remarks

AC Power Cord

Non-Shielded

3.0 m fixed-length

Supplied

USB Cable

Shielded

Max. 3.0 m

Not supplied

HDMI Cable

Shielded

Max. 2.7 m

Not supplied

4.

The use of accessories, transducers and cables other than those specified, with the exception of transducers and

cables sold by Canon sales representative or local Canon dealer as replacement parts for internal components, may

result in increased emissions or decreased immunity of the CR-2.

5.

The CR-2 should not be used adjacent to or stacked with other equipment and if adjacent or stacked use is necessary,

the CR-2 should be observed to verify normal operation in the configuration in which it will be used.