beautypg.com

ROTRONIC HF65 User Manual

Page 29

background image

E-M-HF63_65-V1_11

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

Instruction Manual

Document Type

HygroFlex HF63/HF65 Humidity Temperature
Transmitters: User Guide

Document title

Page 29 of 34

© 2008; Rotronic AG E-M-HF63_65-V1_11

Conformity with standards

HF63

HF65

CE / EMC immunity

EMC Directive 2004/108/EG: EN 61000-6-1: 2001, EN 61000-6-2: 2005
EN 61000-6-3: 2005, EN 61000-6-4: 2001 + A11

Solder type

Lead free (RoHS directive)

Fire protection class

Corresponds to UL94-HB

FDA / GAMP directives

compatible

Environmental limits

HF63

HF65

Storage and transit

-50…+70 °C / -20...+70 °C (models with display), 0…100 %RH, non condensing

Operating limits at electronics

-40 … +60 °C / -10….60 °C (models with display), 0…100 %RH, non condensing

Temperature limits at probe

-100…+150 °C (valid for type D and type 2 – cable probe)

Maximum humidity at sensor

100 %RH up to 80 °C (176 °F)
75 %RH at 100 °C (212 °F)
45 %RH at 125 °C (260 °F)
15 %RH at 150 °C (302 °F)

Maximum air velocity at probe

40 m/s (7,870 ft /min) – depends on dust filter type

Critical environments

Humidity sensor: as per DV04-14.0803.02 - Critical chemicals

This manual is related to the following products: