ORION® Water Endpoints
ORION FIXED NETWORK AND MIGRATABLE ENDPOINTS
This section contains information regarding identification and encoder compatibility for field wiring and activation of ORION
Fixed Network and Migratable endpoints . ORION Fixed Network and Migratable endpoints have a serial number range of
30000000 to 599999999 .
Field Wiring, Encoder Connectivity and Read Resolution
ORION Fixed Network and Migratable endpoints with flying leads are available for connectivity to Badger Meter encoders and
E-Series Ultrasonic meters as well as a number of competitive encoders as shown in chart below .
Endpoints are shipped from the factory pre-programmed and are available in an indoor/outdoor, three-wire configuration for
connection to the encoders listed .
See also
"Endpoint Installation Kits" on page 16
All three wires must be connected to complete an installation . The ORION endpoint connection can be made to either existing
wires from the encoder or directly to the terminal screws of the encoder, depending on the application and manufacturer .
Follow the manufacturer’s installation kit instructions provided with the gel splice or field splice kit you are using .
NOTE:
N
For instructions on using gel connectors for field wiring, see
"Using Gel Caps to Connect an Encoder" on page 14
ORION endpoint wires:
Red
= Power/Clock; Black = Ground;
Green
= Data
Endpoint
Label
Encoder Connectivity
Endpoint Wire Colors
Red
Black
Green
Reading Resolution
ELCD or ENC
Badger Meter HR-E LCD or HR-E encoders, or
E-Series Ultrasonic Meter with High Res output
Enc
oder W
ir
e/T
ermina
tion
Colors
Red
Black
Green
Up to eight (8) most
significant digits
ADE or ENC
Badger Meter ADE or E-Series Ultrasonic Meter with
ADE output
Red
Black
Green
Up to six (6) most
significant digits
RTR
Badger Meter RTR or E-Series Ultrasonic Meter with
RTR output
Red
Black
Green
Up to seven (7) most
significant digits
ADE or ENC
Elster/AMCo ScanCoder or Invision*
Green
Black
Red
Up to eight (8) most
significant digits
C700D
Elster/AMCo C700 Digital*
Red
Black
Not used –
cut green wire
flush with outer
sheath
Up to seven (7) most
significant digits
ADE or ENC
Master Meter Octave Ultrasonic meter
encoder output
Red
Black
Green
Up to eight (8) most
significant digits
ADE or ENC
Metron Hawkeye*
Red
Black
Green
Up to eight (8) most
significant digits
ADE or ENC
Neptune ProRead, E-coder or ARB-V*
Black
Green
Red
Up to eight (8) most
significant digits
ADE or ENC
Sensus ECR II or ICE*
Red
Black
Green
Up to eight (8) most
significant digits
ADE or ENC
Hersey Translator
Due to the customized, factory wire configurations of the Hersey Translator,
the terminal posts may not match the ORION endpoint wire colors . Please
contact Hersey for the terminal post wiring schematic of your encoders to
determine how the posts correspond to ORION endpoint wires .
NOTE:
N
Competitive encoder output is determined by the encoder configuration .
*ORION Fixed Network and Migratable ADE or ENC endpoints are compatible with the encoders noted above with a manufacture date of 2000 or newer as
long as the encoder is programmed into the three-wire output mode for AMR/AMI and has three wires connected to it . Encoder registers that are currently
in two-wire mode of operation require programming by the Utility, including registers that support auto two- or three-wire detection systems that do not
automatically switch to three-wire mode of operation once a compatible endpoint is connected for ORION connectivity .