beautypg.com

Specifications, Specifications 29, 10 specifications – Thermo Fisher Scientific EcoScan Salt 6 User Manual

Page 33

background image

Instruction Manual Salt 6

29

10 SPECIFICATIONS

SPECIFICATIONS DESCRIPTIONS
Percentage (%) Range

0.10 to 5.00 %

Resolution

0.01

%

Accuracy

±1% F.S.

TDS Range

1.0 to 50.0 ppt

Resolution

0.1

ppt

Accuracy

±1% F.S.

Temperature Range

-10.0 to 110.0

°C

Resolution/Accuracy

0.1

°C / ± 0.5 for °C

Cell Constant

1.0

Temperature Compensation

Automatic / Manual (from 0 to 50 °C)

Temperature Coefficient

0.0 to 3.0% / °C

Normalization Temperature

20.0 °C and 25.0 °C (selectable)

Saline Conversion Factor

Non-linear Compensation

Number of calibration points

1 point

HOLD Function

Yes

Auto Power off

20 minutes after last key operation

Inputs

BNC for conductivity and phono jack for temperature

Display

Single Custom LCD

Power Requirements

4 ‘AAA’ Batteries

Battery Life

> 100 hours

Dimension / Weight

Meter: 14 x 7 x 3.5 cm; 200 g