ADVANCED SETUP FUNCTIONS

6.1

Advanced Setup Overview

The advanced setup mode lets you customize your meter’s preferences and
defaults. To enter advanced setup mode:

1)

Make sure that the meter is switched-off.

2) Press

ON

and MODE key simultaneously, holding both keys for 2

seconds. First release ON key first before releasing the MODE key.

3) [StUP] indicator will appear momentarily and [tCo%] will appear

next.

Enter Setup Page.

%

Adjust Temperature Coefficient value from 0.0
to 3.0 %/°C.

Default value is 2.1 %/°C.

C

°

Select Normalization Temperature. Choice of
either 20 °C or 25 °C.

Default value is 25 °C.

To choose measurement parameter, TDS (ppt)
or percentage (%).

User re-set to factory defaults. Choice of “Yes”
or “No”.

Default value is “no”.