Instruction Manual Salt 6

22

Normalization Temperature

You can set the meter to normalize its conductivity measurements to a standard
temperature of either 25 °C or 20 °C.

The default value is 25 °C.

1)

Enter the advanced setup as described in
Section 6.1.

2) Press

▲ or ▼ key until [t.nr °C] appears

on the LCD. Press ENTER key.

3) Press

▲ or ▼ key to select either [25.0

°C

] or [20.0 °C].

4) Press

ENTER

key to select. The meter

will take you back to the menu, [t.nr °C].

5) Press

▲ or ▼ key to move to the next

menu or press CAL to exit to
measurement mode.