Instruction Manual Salt 6

21

6.2

Temperature Coefficient

The temperature coefficient is the amount of change in conductivity per degree
of temperature; it is expressed in percent per °C. Entering the exact
temperature coefficient of your solution lets you accurately compensate
temperature for almost any solution. You can adjust 0.0 to 3.0 % per °C.

Meter default is 2.1% per °C.

1)

Enter the advanced setup as described in
Section 6.1.

%

%

%

HOLD

ENTER

2) Press

▲ or ▼ key until [t.Co %] appears

on the LCD. Press ENTER key.

3) Press

▲ or ▼ key to select a value

between 0.0 to 3.0.

4) Press

ENTER

key to select. The meter will

take you back to the menu, [t.Co %].

5) Press

▲ or ▼ key to move to the next

menu or press CAL to exit to measurement
mode.

6.3