Instruction Manual Salt 6

12

Preparing the Meter for Calibration

Before starting calibration, make sure you are in the correct measurement
mode.

Ensure that you use new Sodium Chloride standard solutions during calibration.
Do not reuse standard solutions as it may be contaminated and affect the
calibration and accuracy of measurements. Use fresh calibration solution each
time you calibrate your meter. Store solutions in a dry and cool environment if
possible.

Always rinse the probe with either deionized water or rinse solution before and
after each calibration/sample measurement to avoid cross-contamination. For
details please refer to Section 7 - Probe Care and Maintenance.

NOTE: These meters are factory set to a temperature

coefficient of 2.1% per °C.
The factory default value for normalization
temperature is 25 °C. If you need to normalize to a
value other than 25 °C, see

Section 6.3 –

Normalization Temperature.