Without atc (manual temperature compensation) – Thermo Fisher Scientific EcoScan Salt 6 User Manual
Page 20

Instruction Manual Salt 6
16
Without ATC (Manual Temperature Compensation)
For manual temperature compensation, simply unplug the probe’s phono jack
(not BNC) from the meter.
To use manual temperature compensation, you need to enter the temperature
value of your process into the meter. This is the value at which the reading will
manually temperature compensates. You can select any temperature between
0 and 50 °C. Default value is 25 °C.
1) Make sure that the phono jack (for
temperature measurement) is disconnected
from the meter. See Section 3.5.
2)
Switch on the meter and if necessary, press
MODE
key to select temperature
measurement mode. See Section 3.7.
3) Press
CAL
key to start temperature
calibration process.
4) The
“CA” will appear momentarily and a
temperature value will starting flashing.
5)
Check the temperature of your sample
using an accurate thermometer. Wait for the
value to stabilize and press ▲ or ▼ key
and adjust the value to the reference
thermometer used.
6) Press
the
ENTER
key. The [CO] indicator
will appear for 1.5 seconds, and the reading
will stop flashing. The temperature
calibration is successfully performed. The
meter returns to measurement mode.
C
°
C
°
C
°
HOLD
ENTER
°C
CAL
5.3