Instruction Manual Salt 6

16

Without ATC (Manual Temperature Compensation)

For manual temperature compensation, simply unplug the probe’s phono jack
(not BNC) from the meter.

To use manual temperature compensation, you need to enter the temperature
value of your process into the meter. This is the value at which the reading will
manually temperature compensates. You can select any temperature between
0 and 50 °C. Default value is 25 °C.

1) Make sure that the phono jack (for

temperature measurement) is disconnected
from the meter. See Section 3.5.

2)

Switch on the meter and if necessary, press
MODE

key to select temperature

measurement mode. See Section 3.7.

3) Press

CAL

key to start temperature

calibration process.

4) The

CA” will appear momentarily and a

temperature value will starting flashing.

5)

Check the temperature of your sample
using an accurate thermometer. Wait for the
value to stabilize and press ▲ or ▼ key
and adjust the value to the reference
thermometer used.

6) Press

the

ENTER

key. The [CO] indicator

will appear for 1.5 seconds, and the reading
will stop flashing. The temperature
calibration is successfully performed. The
meter returns to measurement mode.