8 TROUBLE-SHOOTING

GUIDE

Problem Cause

Solution

Power on but no
display

a) Batteries not in place

b) Batteries not in

correct polarity (+ and
– position).

c) Weak

batteries

a) Check that batteries

are in place and
making good contact.

b) Re-insert

batteries

with correct polarity.

c) Replace batteries.

Unstable
readings

a) Air bubbles in probe.

b) Dirty probe.

c) Probe not deep

enough in sample.

d) External noise pickup

or induction caused
by nearby electric
motor.

e) Broken probe.

a) Tap probe to

remove bubbles.

b) Clean the probe and

re-calibrate.

c)

Make sure sample
entirely covers the
probe sensors.

d) Move or switch off

interfering motor.

e) Replace probe.

Slow response

a) Dirty / Oily probe.

a) Clean probe. See

“Probe Care &
Maintenance”.