beautypg.com

6 maintenance – KROHNE OPTISONIC 7060 EN User Manual

Page 44

background image

OPTISONIC 7060

Electrical & mechanical installation manual 730962.31.00 page 44 of 52


6 Maintenance

6.1 General

The OPTISONIC 7060 does not contain any components that move mechanically. Meter body

and ultrasonic transducers are the only components that come into contact with the

gaseous media. Titanium and high-quality stainless steel ensure that these components are

resistant to corrosion, provided that the device is implemented in accordance with the

relevant specifications. This means that the OPTISONIC 7060 is a low-maintenance system.

Maintenance essentially involves routine checks to determine the plausibility of the

measured and diagnosis values calculated by the system.


It is recommended to record a diagnosis and status log on a regular basis (see software

manual) and compare these values with the initial situation when the system was

commissioned. The operating conditions (gas composition, pressure, temperature, flow

velocity) of the individual logs should be comparable or documented separately and taken

into account when the comparison is evaluated.

6.2 Routine

checks

You can check the front panel of the OPTISONIC 7060 to ensure that the system is

functioning

properly

(see

Section

4.3

).


The routine checks relate to the following values (see also the table below and Section 6).


Velocity of sound


The velocity of sound measured is usually highly stable. Sudden changes in the measured

value can indicate signal detection problems, which can affect propagation time

measurements, or changes in the gas composition. A theoretical velocity of sound value can

be calculated by analysing the gas or recording the pressure and temperature during log

measurement. Implausible measurements can then be indicated by comparing theoretical

and measured velocity of sound values and identifying any marked discrepancies. The

velocity of sound values in the paths should also be approximately equal.


Number of rejected measurements


The number of rejected measurement (% inaccurate measurement) for the measuring

path(s) should be as close to 0 % as possible, although this largely depends on the flow

velocity. With high flow velocities, the figure can be as high as 50% without affecting

accuracy. Marked discrepancies in the values under similar conditions (pressure,

temperature, gas flow rate, gas composition) indicate that changes have been made to the

device or plant (e.g. malfunctions caused by a valve that is not fully open).


Receiving

sensitivity


The receiving sensitivity set by the device largely depends on the process pressure. Under

normal conditions, this value is highly stable. The difference between the ultrasonic

transducers for a measuring path is small, although it can increase with greater velocities.

Significant fluctuations in the receiving sensitivity indicate a low-quality receiving signal. A

significant increase under similar process conditions is normally caused by contamination

on the ultrasonic transducers (for instructions on cleaning them, see the service manual).


Signal-to-noise ratio


These values are typical for the plant and do not change, providing the conditions remain

the same. A reduction in the signal-to-noise ratio with similar reception sensitivity indicates

sources of acoustic interference (e.g. pressure regulator) near the measuring location.