beautypg.com

KROHNE OPTISONIC 7060 EN User Manual

Page 16

background image

OPTISONIC 7060

Electrical & mechanical installation manual 730962.31.00 page 16 of 52

2.5 Self-diagnosis

In the measurement mode, the ratios of sound and path velocities, amplification settings,

and signal-to-noise are continuously monitored. If these parameters deviate from a preset

range, a warning signal is generated. This enables immediate measures to be taken to

prevent possible system malfunctions.


During commissioning or operation, you can adjust the signalling threshold values to suit

individual application requirements. This enables you to create the most effective status

warning

system.


Note

The "Warning" status signal does not affect the device function.


Parameter

Default
threshold
value

Warning
message

Notes

Velocity of
sound

< 5 m/s

Warning SOS
Deviation

This message is produced when the current measured path velocity of
sound deviates from the mean value of the mean velocities of sound
calculated for all the paths by more than the specified threshold value.
The current flow velocity is used as a weighting factor, so that
temperature stratification is disregarded at very low flow velocities. Used
to indicate whether or not the path is measuring the correct propagation
time.
Note
When setting the parameters, take into account plausible conditions for
normal operation (in particular temperature stratification).

< 6 dB

Warning AGC
Deviation

The absolute difference between both path gain factors is evaluated and
must remain below the threshold value.
Important

High flow rates can also increase the difference in gain.

Reception
gain

< 93 dB

Warning AGC
Limit

The absolute value of the reception gain is monitored.
Important

The current receiving sensitivity largely depends on the current process
pressure (inversely proportional in initial approximation, that is, when the
pressure doubles, the required receiving sensitivity is halved).

If one of the alarms is triggered by a path, this can indicate a malfunction
in the ultrasonic transducers, electronics, probe cables or parameter
settings (signal models, standard threshold values).

Signal-to-
noise ratio

< 13 dB

Warning SNR

This alarm is activated when the signal-to-noise ratio is too small.
Reasons for this include interference noise caused by fittings in the
pipeline, valves that are not fully open, sources of noise near the
measuring location, or defective ultrasonic transducers.


Additional signal and system diagnosis functions monitor the accuracy of the measured

values, by checking the plausibility of the ultrasonic signals received and the ultrasound

signal propagation times calculated from them.