Electrical & mechanical installation manual 730962.31.00 page 43 of 52

5.2.3

Pressure testing of a gas pipeline with liquid (water)

For example when a pipeline in which the OPTISONIC 7060 is installed will be pressure tested with
a liquid (water) the following precautions must be taken to prevent the transducers from getting wet:

• The transducers must be removed.

• Blindstops must be fitted in the tranducer positions. A set of blindstops can be ordered at

KROHNE.

For the detailed procedure to remove and refit the transducers, consult the service manual.