beautypg.com

KROHNE OPTISONIC 7060 EN User Manual

Page 12

background image

OPTISONIC 7060

Electrical & mechanical installation manual 730962.31.00 page 12 of 52

2.3.2 Ultrasonic

Transducers

The OPTISONIC 7060 ultrasonic transducers are optimised to suit the system requirements.

The high quality of the transducer parameters provides the basis for accurate and highly

stable propagation time measurements with nanosecond precision. The ultrasonic

transducers are of an intrinsically safe design (class "ia").

2.3.3

Signal Processing Unit (CONVERTER)

The CONVERTER contains all the electrical and electronic components required for controlling the

ultrasonic transducers. It generates transmission signals and uses the received signals to

calculate the measured values. The CONVERTER also contains several interfaces for
communication

with a PC, or standardised process control system.

Event logging

The current volume counter value, fault, warning, and power failure alarms are stored in a

see Section

battery-buffered data memory (FRAM) along with the time of day. On system restart, the

2.7

counter value that was last saved is restored as the start value for the volume counter. The

FRAM back-up provides an unlimited number of writing cycles and protects the saved data

for at least 10 years.


The CONVERTER is supplied with a front panel containing a two-line LCD to display current
measured

values, diagnosis and logbook information (see Fig. 2.4). You can select the values you

want to display using a magnetic pen while the front cover is kept closed (for details on

operation and menu structure see Section 8.2 in the appendix).

Fig. 2.4: Front panel for the OPTISONIC 7060 CONVERTER

1. Selected measured value / device status
2. Current measured value
3. Control area for operation by a magnetic pin