TECHNICAL DATA

7

41

OPTISYS CL 1100

www.krohne.com

01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en

7.2 Technical data

Technical data for the system

OPTISYS CL 1100

Installation conditions

Installation

Wall installation with sample feed

Housing protection

class

IP65, Nema 4

Sample flow

connections

Tube connection DN 6/8

Drill Hole

10 mm / 0.39"

Measuring accuracy

Reference

conditions

Medium: water

Temperature: 20°C / 68°F

Pressure: 1 bar / 14.5 psi (absolute)

Maximum

measuring error

≤ 2% full scale
Temperature: 1.0% full scale

Repeatability

0.2% full scale

Resolution

0.1/0.01 mg/l

Long-term stability 24 hours: tested within accuracy definition

Temperature drift

Tested within accuracy definition

Cable length

variation

Tested within accuracy definition

Operating conditions

Ambient

temperature

-15...+55°C / +5...+131°F

Process

temperature

0…+50°C / +32…+122°F

Max. operating

pressure

4 bar / 58 psi

Min. flow rate

> 30 l/h / 7.93 gal/h

Min. conductivity

> 150 μS/cmTD