Operation – KROHNE OPTISYS CL 1100 EN User Manual
Page 23

OPERATION
5
23
OPTISYS CL 1100
www.krohne.com
01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en
After starting-up the signal converter, the measuring screen appears. This is the standard
screen which is displayed automatically in the normal operating mode. If you are in this mode
and you want to adjust the temperature compensation, you have to perform the following steps:
Step 1: activating the temperature compensation for measurement
Press ^ several time until you reach the measuring mode again. Choose yes to safe and confirm
your selection.
Step 2: Configure/adjust the temperature sensor for measurement
Press > for more than 2.5 seconds, then release the button. You are now on the main menu level. In the
upper line on the right side of the display "A" appears, beneath the main menu quick setup is highlighted.
Press or until the main menu setup is highlighted.
MAIN MENU
A quick setup
B test
> C setup
D service
Press > to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "setup" and "C1"
appears, beneath the submenu process input A is highlighted.
Press > to enter the chosen menu.
You are on the second submenu level. In the submenu parameter
(Cl
2
/ClO
2
/O
3
) is highlighted.
Use or until the submenu temperature is highlighted.
Press > to enter the chosen menu.
You are on the parameter level. In the upper line of the display
"temperature" and "C1.15.1" or "C2.15.1"appears, beneath the option
probe is highlighted.
Use or until the menu item temp. comp. is highlighted.
Press > to enter the chosen menu.
Now you can set up the temperature compensation.
Press or to choose between on and off.
Press ^ to confirm the entered value.
If you have chosen the option linear, you can select the type of
temperature compensation now.
Press or until menu item temp.coefficient is highlighted.
Enter the value and press ^ to confirm the value.
Press or to choose probe.
Press > to enter the chosen menu.
Press or to choose between Pt1000, Pt100 or
manual.
Press ^ to confirm the entered value.
Step 2a: probe Pt100 or Pt1000:
Read the currently measured temperature of the Pt100 / Pt1000 temperature sensor from the
measurement screen and write it down.
Measure the temperature with a reference thermometer and check if it deviates from the temperature
measured by the Pt100 / Pt1000.
Press > for more than 2.5 seconds, then release the button. You are now on the main menu level.
Press or until the main menu setup is highlighted.