Completing the calibration procedure

• As soon as the laboratory value has been determined, go again into the calibration menu and

submenu reference by using or and then press >.

• Enter the laboratory value using or and press ^.

i

The signal converter now matches the "stored value" to the laboratory value and makes a
re-calibration. The slope appears on the display and it is stored in the calibration logbook
(this helps to compare the slope values over the time when re-calibrating).

• Assure that the slope of a new sensor is 50...150 mg/nA, otherwise the sensor has a

malfunction. Press ^ if the slope is okay.

i

The signal converter now asks store value?

• Using or choose yes to store the calibration values or no to discard them.
• Press ^ to confirm.
• Press ^ several times until the questions appears if the configuration should be stored.
• Using or choose yes to store the configuration or no to discard it.

i

You have completed the sensor calibration.

Step 5: switching back to measurement

• Deactivate the function "manual hold" again.

INFORMATION!
If an error occurs during the calibration procedure, the signal converter displays an error
message. Possible causes for an error:

Slope too flat.

Sensor too old.

Wrong electrical connection.