OPERATION
5
31
OPTISYS CL 1100
www.krohne.com
01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en
Completing the calibration procedure
• As soon as the laboratory value has been determined, go again into the calibration menu and
submenu reference by using or and then press >.
• Enter the laboratory value using or and press ^.
i
The signal converter now matches the "stored value" to the laboratory value and makes a
re-calibration. The slope appears on the display and it is stored in the calibration logbook
(this helps to compare the slope values over the time when re-calibrating).
• Assure that the slope of a new sensor is 50...150 mg/nA, otherwise the sensor has a
malfunction. Press ^ if the slope is okay.
i
The signal converter now asks store value?
• Using or choose yes to store the calibration values or no to discard them.
• Press ^ to confirm.
• Press ^ several times until the questions appears if the configuration should be stored.
• Using or choose yes to store the configuration or no to discard it.
i
You have completed the sensor calibration.
Step 5: switching back to measurement
• Deactivate the function "manual hold" again.
INFORMATION!
If an error occurs during the calibration procedure, the signal converter displays an error
message. Possible causes for an error:
•
Slope too flat.
•
Sensor too old.
•
Wrong electrical connection.