beautypg.com

3 calibration, 1 calibration order, 2 temperature compensation – KROHNE OPTISYS CL 1100 EN User Manual

Page 22: Operation

background image

5

OPERATION

22

OPTISYS CL 1100

www.krohne.com

01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en

5.3 Calibration

5.3.1 Calibration order

If you are also measuring the temperature and pH value of the measuring medium, it is essential
to calibrate in the following order, otherwise start with 3. calibrate the sensor:

1. Calibrate temperature measurement as described on the following pages.
2. Calibrate pH measurement (refer to the manual of the pH sensor).
3. Calibrate the sensor as described on the following pages.

5.3.2 Temperature compensation

2 option for temperature compensation during measurement:

• on: linear temperature compensation.
• off: temperature compensation is disabled.

3 option for temperature compensation during calibration:

• automatic: the signal converter will automatically compensate temperature influence using

the information of a Pt100 or Pt1000 temperature sensor.

• manual: the signal converter will compensate temperature influence using a manually

entered value; this option only makes sense if the temperature of the measured medium is
quite constant.

• off: temperature compensation is disabled.

The menu for the type of linear temperature probes offer the following options:

• Pt1000: choose this option if there is an external Pt1000 temperature sensor connected to the

signal converter.

• Pt100: choose this option if the pH sensor has an integrated Pt100 temperature

measurement or if there is an external Pt100 temperature sensor connected to the signal
converter.

• manual: choose this option if there is no internal or external temperature sensor connected

to the signal converter.

INFORMATION!

The measurement is temperature dependent due to the linear temperature compensation.
The linear temperature compensation compensates the influence of the ion velocity.

Please note that the temperature sensor should always record the temperature where the
measuring electrodes are exposed.

If you activate the temperature c the temperature coefficient can be changed. The presetting
is based on drinking water: 3%/

°

K. You have to change the presetting if the medium is not

drinking water. The coefficient depends on the hardness and the ion velocity of the medium.

If you choose no compensation, the measured Cl

2

, ClO

2

or O

3

concentration will most likely

deviate from the actual Cl

2

, ClO

2

or O

3

concentration. The reason is that the Cl

2

, ClO

2

or O

3

concentration of a specific medium varies depending on the temperature of the medium.