Operation – KROHNE OPTISYS CL 1100 EN User Manual

Page 28

background image

5

OPERATION

28

OPTISYS CL 1100

www.krohne.com

01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en

Step 1: activating the manual hold function

You have activated the manual hold function. To perform the next step and prepare the
calibration procedure you have to return to the measuring mode.
Press ^ until you reach the measuring mode again.

Step 2: preparing the calibration procedure

• If you calibrate a new sensor, make sure that the sensor is correctly connected to the signal

converter.

• Check the sensor for damages or dirt deposits.
• During the calibration procedure, you will have to take a sample. Provide a suitable sample

vessel.

Press > for more than 2.5 seconds, then release the button.

You are on the main menu level. In the upper line on the right side of the display "A" appears, beneath the

main menu quick setup is highlighted.

MAIN MENU

> A quick setup

B test

C setup

D service

Press > to enter the chosen menu.

You are on the first submenu level. In the upper line of the display "quick

setup" and "A1" appears, beneath the submenu language is highlighted.

Press or until the submenu manual hold is highlighted.

Press > to enter the chosen menu.

You are on the parameter level. The option off is highlighted.

Press or to choose the option on.

Press ^ to confirm your selection.