Operation – KROHNE OPTISYS CL 1100 EN User Manual
Page 28
5
OPERATION
28
OPTISYS CL 1100
www.krohne.com
01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en
Step 1: activating the manual hold function
You have activated the manual hold function. To perform the next step and prepare the
calibration procedure you have to return to the measuring mode.
Press ^ until you reach the measuring mode again.
Step 2: preparing the calibration procedure
• If you calibrate a new sensor, make sure that the sensor is correctly connected to the signal
converter.
• Check the sensor for damages or dirt deposits.
• During the calibration procedure, you will have to take a sample. Provide a suitable sample
vessel.
Press > for more than 2.5 seconds, then release the button.
You are on the main menu level. In the upper line on the right side of the display "A" appears, beneath the
main menu quick setup is highlighted.
MAIN MENU
> A quick setup
B test
C setup
D service
Press > to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "quick
setup" and "A1" appears, beneath the submenu language is highlighted.
Press or until the submenu manual hold is highlighted.
Press > to enter the chosen menu.
You are on the parameter level. The option off is highlighted.
Press or to choose the option on.
Press ^ to confirm your selection.