Operation – KROHNE OPTISYS CL 1100 EN User Manual

Page 24

background image

5

OPERATION

24

OPTISYS CL 1100

www.krohne.com

01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en

Press ^ to confirm the entered value. The temperature sensor has been adjusted.

If necessary, enter the temperature correction in Kelvin so that the signal converter shows the
same temperature as the reference thermometer.

MAIN MENU

A quick setup

B test

> C setup

D service

Press > to enter the chosen menu.

You are on the first submenu level. In the upper line of the display "SETUP"

and "C1" appears, beneath the submenu process input A is highlighted.

Press > to enter the chosen menu process input A.

Press or until the main menu temperature is highlighted.

Press > to enter the chosen menu. The submenu probe Pt100 / 1000 is

highlighted.

Press or until the submenu correction is highlighted.

Press > to enter the chosen menu.

INFORMATION!
A separate temperature sensor has to be connected to "Pos.A" and configurated on process
input A.