OPERATION
5
21
OPTISYS CL 1100
www.krohne.com
01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en
C1.13.4
clean
Start cleaning procedure, manually.
C1.15
C2.15
temperature
Menu for temperature measurement. Available for sensor 1 and sensor 2.
C1.15.1
C2.15.1
probe
Options:
•
manual: used if no internal or external temperature sensor is connected to
the signal converter
•
Pt100: used if the pH sensor has an integrated Pt temperature
measurement or if an external Pt100 temperature sensor is connected to
the signal converter
•
Pt1000: used if an external Pt1000 temperature sensor is connected to the
signal converter
C1.15.2
C2.15.2
manual
Only available if C1.15.1 or C2.15.1 is set to "manual".
C1.15.3
C2.15.3
correction
Offset correction for temperature measurement. Not available if C1.15.1 or
C2.15.1 is set to "manual".
C1.15.4
C2.15.4
limitation
Measuring ranges for temperature measurement.
C1.15.5
C2.15.5
temp. comp.
Menu for activating the temperature compensation parameters.
For detailed information refer to
Temperature compensation
on page 22.
Options:
•
on: linear temperature compensation.
•
off: temperature compensation is disabled.
C1.18
product calibration
For detailed information refer to
Calibration
on page 22.
C1.18.1
temp.comp
Menu for activating the temperature compensation parameters for the
calibration.
Options:
•
off: temperature measurement is disabled
•
manual: temperature value has to be entered manually
•
automatic: temperature measurement is performed as configured
C1.18.2
temperature
Menu for setup of manual temperature measurement
C1.18.3
temp. coefficient
Temperature coefficient for manual temperature compensation during
calibration.
C1.18.4
pH compensation
Options:
•
off: pH measurement is disabled
•
manual: pH value has to be entered manually
•
automatic: pH measurement is performed as configured
C1.18.5
man. pH compensation
If in menu C1.18.4 the pH compensation was set to manual, here the desired
pH value is entered manually.
C1.18.6
start calibration
Start calibration procedure.
C1.18.7
stored value
View stored value of calibration.
C1.18.8
reference
Enter the reference value of the probe in mg/L.
CAUTION!
If you choose for measurement the temperature compensation "linear" than choose between
"automatic" or "manual" for the temperature compensation during calibration. If you choose for
measurement the temperature compensation "off" than choose also "off" for the temperature
compensation during calibration.
Level
Designation / function
Settings / descriptions