OPERATION

5

21

OPTISYS CL 1100

www.krohne.com

01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en

C1.13.4

clean

Start cleaning procedure, manually.

C1.15

C2.15

temperature

Menu for temperature measurement. Available for sensor 1 and sensor 2.

C1.15.1

C2.15.1

probe

Options:

manual: used if no internal or external temperature sensor is connected to
the signal converter

Pt100: used if the pH sensor has an integrated Pt temperature
measurement or if an external Pt100 temperature sensor is connected to
the signal converter

Pt1000: used if an external Pt1000 temperature sensor is connected to the
signal converter

C1.15.2

C2.15.2

manual

Only available if C1.15.1 or C2.15.1 is set to "manual".

C1.15.3

C2.15.3

correction

Offset correction for temperature measurement. Not available if C1.15.1 or

C2.15.1 is set to "manual".

C1.15.4

C2.15.4

limitation

Measuring ranges for temperature measurement.

C1.15.5

C2.15.5

temp. comp.

Menu for activating the temperature compensation parameters.

For detailed information refer to

Temperature compensation

on page 22.

Options:

on: linear temperature compensation.

off: temperature compensation is disabled.

C1.18

product calibration

For detailed information refer to

Calibration

on page 22.

C1.18.1

temp.comp

Menu for activating the temperature compensation parameters for the

calibration.

Options:

off: temperature measurement is disabled

manual: temperature value has to be entered manually

automatic: temperature measurement is performed as configured

C1.18.2

temperature

Menu for setup of manual temperature measurement

C1.18.3

temp. coefficient

Temperature coefficient for manual temperature compensation during

calibration.

C1.18.4

pH compensation

Options:

off: pH measurement is disabled

manual: pH value has to be entered manually

automatic: pH measurement is performed as configured

C1.18.5

man. pH compensation

If in menu C1.18.4 the pH compensation was set to manual, here the desired

pH value is entered manually.

C1.18.6

start calibration

Start calibration procedure.

C1.18.7

stored value

View stored value of calibration.

C1.18.8

reference

Enter the reference value of the probe in mg/L.

CAUTION!
If you choose for measurement the temperature compensation "linear" than choose between
"automatic" or "manual" for the temperature compensation during calibration. If you choose for
measurement the temperature compensation "off" than choose also "off" for the temperature
compensation during calibration.

Level

Designation / function

Settings / descriptions