2 menu b, test, B1, sim.process inp.a b2, sim.process inp.b, B4, actual values – KROHNE OPTISYS CL 1100 EN User Manual
Page 19: Operation
OPERATION
5
19
OPTISYS CL 1100
www.krohne.com
01/2013 - 4002492301 - MA OPTISYS CL 1100 R02 en
5.2.2 Menu B, test
The procedure to start the simulation process is the same for all functions:
• Choose the function with the help of ↓ or ↑ and press ^.
i
You see the two options "set value" (opens the editor to enter the simulation value) and
"break" (exits the menu without simulation).
• Choose the desired option with the help of ↑ or ↓ and press ^.
i
If you chose "set value", the device asks "start simulation?" and offers the options "no"
(exits the menu without simulation) or "yes" (starts the simulation finally).
• Choose the desired option with the help of ↑ or ↓ and press ^.
i
If you chose "yes", the simulation starts.
B1, sim.process inp.A
B2, sim.process inp.B
B4, actual values
INFORMATION!
Note that the appearance of some sub-menus depends on the hardware setting and the used
sensor(s). Also only the sensor relevant menus and sub-menus are shown here in detail. For all
other menu functions refer to the converter manual.
Level
Designation / function
Settings / descriptions
B1.1
B2.1
temperature
In this menu the temperature can be simulated.
B1.4
B2.4
conc.absolute
In this menu the concentration can be simulated.
B2.5
REDOX/ORP
In this menu the relative concentration of REDOX/ORP can be simulated. For
2 channel version only; refer to pH/ORP sensor manual for further
information. ONLY if switched to ORP.
B2.7
pH
In this menu the relative concentration of pH can be simulated. For 2
channel version only; refer to pH/ORP sensor manual for further
information.
Level
Designation / function
Settings / descriptions
This menu groups several functions which allow to display the corresponding actual reading. The shown measurements
are depending on the device configuration.
B4.1
operating hours
This menu shows the operating time of the devices in hours.
B4.2
process input A
In this menu the measurements from process input A can be read.
B4.3
process input B
In this menu the measurements from process input B can be read.
For 2 channel version only; refer to pH/ORP sensor manual for further
information.