beautypg.com

Jenway 73 Series PC software User Manual

Page 58

background image

58

To continue click Yes, to cancel click No or Cancel.

If the wavelength is changed then the zero calibration must be performed again before a sample can

be measured. Alternatively a new time scan & kinetics tab can be opened and the required settings

can be entered.

7.3

SAMPLE MEASUREMENT

Insert the cuvette containing the

sample into the sample chamber and

close the instrument lid.

Click on the Scan button; the

instrument initiates the scan and the

live spectrum scan is displayed on

the screen. The results are also

plotted in a table. To stop the scan

before completion click on the Stop

button. To restart the scan click on

the Repeat button – the scan is

restarted from the beginning.

If the same sample needs to be measured more than once click on the Auto Log button. The

instrument will perform the required number of readings with the specified time interval between each

measurement. Unlike with photometrics, concentration and quantitation the auto logged results will

not be displayed on the same screen. For each repeat measurement required a new tab is

automatically opened displaying the new kinetics run and results

.

Auto logging can be stopped before

all the measurements have been completed by clicking on the Stop button. To recommence Auto

logging click on the Auto Log button again.

If the settings menu is re-entered following sample measurement it is not possible to change the

wavelength, run time, time interval or start on level. A new time scan kinetics tab can be opened and

the new settings can be entered.