33

5.1.6

Auto Logging

Auto

logging

allows

repeat

measurements of the same

sample to be performed with a

set time interval between each

measurement. Use the arrow

icons to increase or decrease the

time interval between sample

measurements and use the

arrow icons to select the number

of readings required. If auto

logging is not required the values

do not need to be adjusted in this

screen.

5.2

CALIBRATION

In the spectrum measurement mode the calibration is a baseline scan which is performed across the

selected wavelength range at the selected scan interval.

Insert a cuvette containing the blank solution

into the sample chamber and close the instrument lid.

Click on the Baseline button to begin

the baseline scan. The calibration

status will show the progress of the

baseline scan. Once the calibration

is completed the Scan button

becomes active and the sample can

be measured.

If the settings menu is re-entered following the zero calibration it is possible to change the

wavelength

and scan interval. However a warning will be displayed to warn that critical settings are being

changed after the baseline has already been performed.