beautypg.com

Jenway 7315 Manual Italian User Manual

Page 74

background image

err 5

Light Saturation not Found

This error indicates that the peak light hasn’t been found at zero.

The most likely causes of this error are:

Fatal

1. Lamp failure.

2. Deterioating lamp signal.

3. Sample or cuvette in the sample holder.

Solution: Ensure that the sample holder is empty. Restart the unit, if the

problem persists contact a service technican.

err 7

USB Module Failure

This error indicates the instrument is unable to detect a response from

the UBS module.

Warning

The most likely causes of this error are:

1. USB port failure.

Solution: press the key adjacent to the tick icon to carry on using the

instrument. Restart the unit, if the problem persists contact a service

technician.

err 8

Unable to find Vane on turret

This error indicates that the vane position zero on the turret cannot

be found.

Warning

The most likely causes for this error are:

1. The turret carousel has been removed and not replaced.

Redo

Solution: Check if the turret is in the sample chamber and inserted

properly. Press the key adjacent to the

back icon to try rechecking.

err 9

over temperature

This error indicates that the thermal switch has cut out.

The most likely causes for this error are:

Fatal

1. Failure of the fan

2. Thermal switch not connected.

Solution: restart the unit, if the problem persists contact a service

technician.

74

This manual is related to the following products: