11.2.3.1 Automatic 8 cell turret
Take the 8 cell turret base plate. Connect the power
supply in the bottom of the sample chamber to the
connector on the underside of the baseplate. Place the
base plate in the sample chamber. Replace screws 1 to 4.
Take the 8 cell carousel and place on top of the motor,
taking care to align the three ball bearings with the
grooves on the motor shaft. Gently push the carousel
down onto the motor shaft until it is located into place.
Gently rotate the carousel until there is some resistance.
The carousel is now in the correct position.
If the fitting is too tight use a small screw driver to loosen
the ball bearings before pushing the carousel down onto
the shaft.
11.2.3.2 Peltier
For this accessory as well as removing the passive
accessory base plate, the front panel of the instrument
must also be removed. Loosen screws 5 and 6 until the
front panel can be lifted out in the forwards direction.
Take the peltier base plate. Connect the power supply
in the bottom of the sample chamber to the connector
on the underside of the base plate. Place the base plate
in the sample chamber. Replace screws 1 to 4. Take the
peltier front panel and slot into place before retightening
screws 5 and 6.
When the accessory is fitted the instrument will look like
this.