Jenway 7315 Manual Italian User Manual
Page 52
52
10.1.2
SAVinG MetHoDS to USB MeMoRY StiCK
In the expanded operating menu press the key adjacent
to the
method icon to open the method selection
menu. Select the method to save to USB memory stick
by pressing the key adjacent to the method location.
Use the keys below the
arrow icons to move page up or
down. Hold the key below the
create method icon for
2 seconds to open the save to UBS memory stick menu.
Either the single method selected or all 48 methods can
be saved to USB memory stick. To save a single method
press the key adjacent to the
single method save icon.
To save all 48 methods press the key adjacent to the
1
- 48 method save icon. Press the key adjacent to the
tick icon to confirm saving the method(s) to USB memory
stick. Press the key adjacent to the
cross icon to cancel
saving the method(s) to USB memory stick and return to
the method selection screen.
Confirmation will be needed to save the methods to USB
memory stick. Press the key adjacent to the
tick icon to
confirm saving the method(s). Press the key adjacent
to the
cross icon to cancel and return to the option of
saving one or all of the methods to USB memory stick.
10.2
oPeninG MetHoDS
Methods can be opened from the internal memory of the instrument or from a USB memory stick.
Before a method can be opened from a USB memory stick it must be saved to the internal memory of
the instrument first.
10.2.1
oPeninG MetHoDS FRoM inteRnAL MeMoRY
In the expanded operating menu press the key adjacent
to the
method icon to open the method selection menu
and use the keys below the
arrow icons to move page
up or down. Select the method to open by pressing the
key adjacent to the method location. Press the key below
the
open method icon.
The instrument will return to the current measurement
mode and the method name will be displayed at the top
of the expanded operating menu.
If a method hasn’t been selected it is still possible to calibrate the instrument and take measurements.
The instrument will operate in a generic method, 0, where all the current operating settings will be
stored. There are separate method 0’s for each measurement mode.