Jenway 7315 Manual Italian User Manual
Page 24
24
The time set will begin to count down when there is
no lamp activity. When the count down is complete the
lamp and the fan will be turned off and the
lamp save
icon is shown in all the measurement modes. To bring
the instrument out of the lamp save in order to perform
a measurement press the key below the
lamp save icon.
The lamp and fan will be turned back on and the lamp
will begin to warm up.
The
lamp cold icon is displayed adjacent to the calibrate
to zero icon in the measurement mode. The time needed
for the lamp to warm up is five minutes.
To view the remaining warm up time press the key below
the
lamp cold icon. Calibration and measurements can
be performed whilst the lamp is warming up but these
results may not be accurate.
Once the warm up time of five minutes is complete the
lamp cold icon disappears.