9.2.1.8 Selecting a Wavelength

To adjust the wavelength press the key adjacent to the
wavelength icon. This will open the number entry
screen. Use the keys at the bottom of the screen to select
the digit to be changed and the keys adjacent to the
arrow icons to increase or decrease the number. Once
the required wavelength has been entered press the key
adjacent to the

tick icon to save the changes and return

to the settings menu screen.

9.2.1.9 Setting the Kinetics Measurement time

This function enables the measurement time for the
kinetics scan to be set. To access this function press the
key below the time displayed on the far right of the x-axis;
this will open a number entry screen. Use the keys at the
bottom of the screen to select the digit to be changed
and use the keys adjacent to the

arrow icons to increase

or decrease the number. Once the required measurement
time in seconds has been entered press the key adjacent
to the

tick icon to save and return to the settings menu.

9.3

CALiBRAtion

The calibration must be performed at the same
wavelength at which the sample will be measured. Insert
a cuvette containing the blank solution into the sample
chamber and close the instrument lid. Press the key below
the

calibrate to zero absorbance icon. The instrument

will calibrate to 0.00 absorbance.

Once the calibration is complete the

start kinetics measurement icon will be displayed and a sample

can be measured. If the wavelength is adjusted before the sample is measured the

start kinetics

measurements icon will disappear and the instrument must be calibrated again at the new wavelength.

9.4

SAMPLe MeASUReMent

Insert a cuvette containing the sample to be analysed
into the sample chamber and press the key below the
start kinetics measurements icon. If a start on level or
lag time has not been set the instrument begins taking
a photometric reading every second. As each reading is
taken it is plotted on the graph.

If a lag time has been set a progress bar will be displayed
showing the lag time counting down. Once the lag time
is complete the instrument will begin taking readings.

4

0

0

0

0

2

0

0

3.000

ABS

-5.000

0.00

10.00

5.0

s

09:02

3.000

ABS

-5.000

0.00

10.00

5.0

s

09:02

09:02

2.500

ABS

-0.500

1.000

350

550

450nm

0

100

This manual is related to the following products: