A depth reference applies an offset equal to the distance from the
pressure sensor to a desired location, such as the bottom of the
probe, so that the depth reading is reported from the desired
location, rather than from the location of the pressure sensor.

1

Pressure sensor

2

Example of a Depth reference setting

6. When the calibration is stable, tap the Accept button.

Calibrate the pH Sensor

1. Tap the Calibration icon

to access a list of sensors that are available for

calibration.

2. Tap pH Sensor.

3. Select the method by which to calibrate the sensor. This example demonstrates a

two-point calibration. Tap 2-Point Calibration.