Setting the depth reference – In-Situ smarTROLL Multiparameter Handheld Operators Manual User Manual

Page 45

background image

45

Do not perform a "Zero in Air" calibration if a Depth reference is
already set because it will result in a faulty calibration. To clear a
Depth reference, tap Restore Defaults.

4. Ensure that the pressure sensor is exposed to air and is not submerged in liquid.

5. Tap the Start button.

6. When the calibration is stable, tap the Accept button.

Setting the Depth Reference

1. Tap the Calibration icon

to access a list of sensors that are available for

calibration.

2. Tap Depth Sensor.

3. Tap Depth Reference.