beautypg.com

In-Situ smarTROLL Multiparameter Handheld Operators Manual User Manual

Page 38

background image

38

The dissolved oxygen sensor cannot be calibrated using the
Quick-Cal procedure.

4. Tap OK.

5. Fill the calibration cup to the fill line with Quick-Cal solution.

6. Place the instrument into the calibration cup, and tap Start.

7. When the calibration is stable, tap the Accept button.

This manual is related to the following products: