beautypg.com

GW Instek LCR-914 User Manual

Page 28

background image

LCR-914/915/916 User Manual

26

(7) DCR

Range

Resolution

Accuracy

200.00Ω

0.01

Ω

± (0.2% + 5d)

2.0000k

Ω

0.0001k

Ω

± (0.2% + 2d)

20.000k

Ω

0.001k

Ω

200.00k

Ω

0.01k

Ω

2.0000M

Ω

0.0001M

Ω

20.000M

Ω

0.001M

Ω

± (0.5% + 2d)

200.00M

Ω

0.01M

Ω

± (1.0% + 2d)

[1]

[1] < 50dgt rolling.

Input Protection:
30V

DC

or 30V

AC

rms

Minimum Resolution: 0.01

Ω in the 200.00Ω range.

Measuring Time: 2 seconds

(8) D & Q

Definition Q=1/D
Range:
2.000 ~ 2000
Minimum Resolution: 0.001 in the 2.000 range.
Accuracy: Accuracy of Main Reading x (1+D)
Input Protection: 30V

DC

or 30V

AC

rms

(9) ESR

The specification for ESR is the same as Resistance.

(10)

θ

Range

Resolution

Accuracy

-90.0° ~ 90.0°

0.1°

± (0.2% + 5d)


Input Protection:
30V

DC

or 30V

AC

rms





This manual is related to the following products: