LCR-914/915/916 User Manual

17

When “Srt” appears on the sub display, short the terminals or the
SMD test probe, and then press the CAL button to start the short
calibration. About 30 seconds later, the result of the short
calibration appears on the main display. If the result is

“pass”,

press the CAL button to complete the calibration. If the result is
“fail”, press the CAL button to exit the function.