beautypg.com

GW Instek LCR-914 User Manual

Page 25

background image

LCR-914/915/916 User Manual

23

(4) Inductance

Frequency

Range

Accuracy

100Hz
120Hz

20.000mH

± (0.5% + 5d)

[2]

200.00mH

± (0.2% + 2d)

2000.0mH

20.000H
200.00H
2000.0H

± (0.5% + 2d)

20.000kH

± (1.0% + 2d)

[2]

1kHz

2000.0uH

± (0.5% + 5d)

[2]

20.000mH

± (0.2% + 2d)

200.00mH
2000.0mH

20.000H
200.00H

± (0.5% + 2d)

2000.0H

± (1.0% + 2d)

[2]

10kHz

[1]

200.00uH

± (0.5% + 5d)

[2]

2000.0uH

± (0.2% + 2d)

20.000mH
200.00mH
2000.0mH

± (2.0% + 2d)

20.000H

± (5.0% + 2d)

100kHz

[3]

20.000uH

± (0.5% + 5d)

[2]

200.00uH

± (0.2% + 2d)

2000.0uH

20.000mH

± (2.0% + 2d)

200.00mH

± (5.0% + 2d)

[1] The 10kHz test frequency is only for the LCR-915/916.
[2] The measuring time is 2 seconds.
[3] The 100kHz test frequency is only for the LCR-916.

Input Protection: 30V

DC

or 30V

AC

rms

Minimum Resolution: 0.001uH in the 20.000uH range.
Measuring Time: 800ms

Note: If D > 0.1, the accuracy should be multiplied by

2

1 D