GW Instek LCR-914 User Manual
Page 26

LCR-914/915/916 User Manual
24
(5) Capacitance
Frequency
Range
Accuracy
100Hz
120Hz
2000.0pF
± (0.5% + 5d)
[2]
20.000nF
± (0.2% + 2d)
200.00nF
2000.0nF
20.000uF
200.00uF
± (0.5% + 2d)
2000.0uF
± (1.0% + 2d)
20.000mF
± (2.0% + 2d)
[2]
1kHz
2000.0pF
± (0.5% + 5d)
[2]
20.000nF
± (0.2% + 2d)
200.00nF
2000.0nF
20.000uF
± (0.5% + 2d)
200.00uF
± (1.0% + 2d)
2000.0uF
± (2.0% + 2d)
[2]
10kHz
[1]
200.00pF
± (0.5% + 5d)
[2]
2000.0pF
± (0.2% + 2d)
20.000nF
200.00nF
2000.0nF
20.000uF
± (2.0% + 2d)
200.00uF
± (5.0% + 2d)
[2]
100kHz
[3]
20.000pF
± (0.5% + 20d)
[2]
200.00pF
± (0.2% + 2d)
2000.0pF
20.000nF
200.00nF
2000.0nF
± (2.0% + 2d)
20.000uF
± (5.0% + 2d)
[2]
[1] The 10kHz test frequency is only for the LCR-915/916.
[2] The measuring time is 2 seconds.
[3] The 100kHz test frequency is only for the LCR-916.
Input Protection: 30V
DC
or 30V
AC
rms
Minimum Resolution: 0.001pF in the 20.000pF range.
Measuring Time: 800ms
Note: If D > 0.1, the accuracy should be multiplied by
2
1 D