LCR-914/915/916 User Manual

22

Shock Vibration: Sinusoidal vibration per MIL-T- 28800E
(5 ~ 55 Hz, 3g maximum).
Drop Protection: 4 feet drop to hardwood on concrete floor.
Indoor Use.


Electrical Specifications

(1) Test Frequency

Range

Resolution

Accuracy

100.00 Hz

0.01 Hz

± 0.01%

120.00 Hz

0.01 Hz

± 0.01%

1.0000 kHz

0.1 Hz

± 0.01%

10.000 kHz

[1]

1 Hz

± 0.01%

100.00 kHz

[2]

10 Hz

± 0.01%

[1] The 10kHz test frequency is only for the LCR-915/916.
[2] The 100kHz test frequency is only for the LCR-916.

(2) Test Signal

AC Signal Level: 600mVrms
AC Signal Accuracy:
±10%
DC Bias Level: 1V
DC Bias Accuracy:
±10%

(3) Test Cable

Model

Length

Bandwidth

Type

SMD Test Probe

60cm

1MHz

5-Wire

4-Wire Test Probe

60cm

1MHz

5-Wire

Alligator Clip Set

15cm

1kHz

2-Wire


Accuracy:
±(A x B)(% of reading)
A: Basic Accuracy as specified by
B: Test Cable Accuracy
B(%) = 1 + (L x F x T)
L(m): Cable length
F(MHz): Test frequency
T: Cable type. If the cable is 5-

wire type, then “T” is 40, else “T”

is 4,000.

When measuring by basic accuracy, the following
conditions must be met:

1.

Ambient temperature: 23˚C ± 5˚C < 80% RH.

2.

Test cable length: 0m

3.

Open and short calibration has been performed

4.

D ≤ 0.1 for C, L measurements; Q ≤ 0.1 for R
measurements.

See the operation manual for additional details.