PSP SERIES

USER MANUAL

6

The power supply outputs and connecting leads, sockets and terminals

must be protected from being touched directly. In addition, the leads

used must be sufficiently insulated and voltage-proof and the contact

points safe from being touched (safety sockets).

Use of bare metal leads and contacts should be avoided. All such items

are to be covered by suitable, nonflammable insulating material or

other measures taken and therefore protected from being touched

directly. The electrically conducting parts of the connected load must

also be appropriately protected from being touched directly.

If the instrument is assumed that safe operation is no longer possible

according to the following phenomena, then the unit must be switched

off and protected against unintentional operation:

- the unit shows visible signs of damage,

- the unit no longer functions and

- after prolonged storage under unfavourable conditions or

- after severe transportation stress.

Do switch on the switched power supply immediately when it has been

brought from a cold environment into a warm room until it comes to

room temperature as under adverse conditions, the resultant

condensation can destroy the unit.

PSP SERIES

USER MANUAL

7

3. Technical Specifications

Model

PSP-603

PSP-405

PSP-2010

Operating voltage

115/230 VAC ±15%

Power frequency

50/60 Hz

Power consumption

approx. 420VA max.

Power output

200W max.

Output voltage

0~60VDC, 20mV

resolution

0~40VDC,

10mV resolution

0~20VDC, 10mV

resolution

Program Accuracy

±0.05%±4 digits

±0.05%±3 digits

±0.05%±3 digits

Output Current

0~3.5A

2mA resolution

0~5A

2mA resolution

0~10A

5mA resolution

Program Accuracy

±0.1%±5 digits

±0.1%±5 digits

±0.3%±10 digits

Voltage Load Regulation

≦ 10 mV

Current Load Regulation

≦ 5 mA

Voltage Line Regulation.

≦ 0.05%

Current Line Regulation

≦ 0.05%

Ripple Voltage

≦ 20 mV rms

Ripple Current

≦ 10 mArms

Readback Resolution
(Meter)

20mV 2mA

10mV 2mA

10mV 5mA

Rise Time

≦150ms (≦95% rating load)

Response
Time

Fall Time

≦150ms (≧10% rating load)

Recovery Time

30ms (50% Load Change, Minimum load 0.5A)

Readback Accuracy (Meter)

Voltage

±0.05%±4 digits

±0.05%±3 digits

±0.05%±3 digits

Current

±0.1%±5 digits

±0.1%±5 digits

±0.3%±10 digits

Digital Display

Multi-line LCD with background lighting

AC fuse

Slow-blow T6.3A/250V for 115V, T3.15A/250V for 230V

Command Process Time

250ms

Interface (Standard)

RS-232C

Weight

Approx. 4 kgs

Dimensions
(W × H × L)

Approx. 225Ч100Ч305 m/m

(without stand and power cable)