USER MANUAL

32

Example:

The present voltage limit is at 30V, and the wheel knob status is at
normal, the SU- message is sent to the power supply, the voltage
limit of which will become 29V.

*SI+

Function:

Add one unit to the present current limit setting.

Syntax:

SI+

HEX = 53 49 2B 0D

Explain:

When the message of SI+ is sent to the power supply from
computer, the power supply will add one unit to the present current
limit setting immediately.

Example:

The present current limit is at 3.00A, and the wheel knob status is at
normal, the SI+ message is sent to the power supply, the current
limit of which will become 3.10A.

*SI-

Function:

Subtract one unit from the present current limit setting.

Syntax:

SI-

HEX = 53 49 2D 0D

Explain:

When the message of SI- is sent to the power supply from
computer, the power supply will subtract one unit from the present
current limit setting immediately.

PSP SERIES

USER MANUAL

33

Example:

The present current limit is at 3.00A, and the wheel knob status is at
normal, the SI- message is sent to the power supply, the current
limit of which will become 2.90A.

*SP+

Function:

Add one unit to the present load limit setting.

Syntax:

SP+

HEX = 53 50 2B 0D

Explain:

When the message of SP+ is sent to the power supply from
computer, the power supply will add one unit to the present load limit
setting immediately.

Example:

The present load limit is at 100W, and the wheel knob status is at
normal, the SP+ message is sent to the power supply, the load
limit of which will become 101W.

*SP-

Function:

Subtract one unit from the present load limit setting.

Syntax:

SP-

HEX = 53 50 2D 0D

Explain:

When the message of SP- is sent to the power supply from
computer, the power supply will subtract one unit from the present
load limit setting immediately.