GW Instek PSP Series User Manual
Page 10
PSP SERIES
USER MANUAL
14
the "V LIMIT " key is now pressed for longer than 2s,
the voltage upper limit will again be set at "rating
voltage". To complete the input, press the "ENTER"
key. An inadvertently wrong-set limit setting can be
erased/reset with the " CE " key. As a result of this, the
previous set value will be indicated and the setting
menu exited. The "U-const" will no longer flash.
During operation (output On), the voltage output can
also be adjusted up to the set limit.
I LIMIT Current limit
Press the " I LIMIT" key in the "LIMITS" field until the
symbol "I-const" flashes. The current limit can now be
adjusted to 1mA, 2mA, 10mA ("fine") or 100mA
("norm", coarse) steps by using the encoder wheel. If
the "I LIMIT" key is now pressed for longer than 2s, the
current upper limit will again be set at "rating current".
To complete the input, press the "ENTER" key. An
inadvertently wrong-set limit setting can be erased/reset
with the " CE " key. As a result of this, the previous set
value will be indicated and the setting menu exited. The
"I-const" will no longer flash.
P LIMIT Power limit
Press the "P LIMIT" key in the "LIMITS" field until the
symbol "P-const" flashes. The power limit can now be
adjusted to 1 W steps by using the encoder wheel. If the
"P LIMIT" key is now pressed for longer than 2s, the
PSP SERIES
USER MANUAL
15
power upper limit will again be set at "200 W". To
complete the input, press the "ENTER" key. An
inadvertently wrong-set limit setting can be erased/reset
with the " CE " key. As a result of this, the previous set
value will be indicated and the setting menu exited. The
"P-const" will no longer flash.
Attention: The current "I LIMIT" is adjusted
independently of its setting by the maximum output
power setting. The voltage upper limit does not
change.
C) "KEY INPUT" field settings
V SET Using the "V SET" key, the output voltage can be adjusted
up to the preset upper limit directly. For this the "V SET"
key must be pressed and held and the voltage changed
with the encoder wheel. If the "fine" F key is selected,
the changes occur in 10 mV, and 20mV steps. If
"norm" N key is selected, the changes occur in 1V
steps.
ENTER Each input via the "LIMITS" keypad is completed with the
"ENTER" key.
CE Any values incorrectly entered via the "LIMITS" keypad
can be reset with the " CE " key.
N The size of each setting step can be changed by using the
" N " ("Normal") key. In connection with the " F " key, so-
called mutual locking is achieved.
F If the " N " key is pressed, the "fine" symbol disappears