Model PI-700

PI-700 Instruction Manual

Rev. 2.5

Page 40 of 54

 Verify adequate Sensor Life.

 Clean or replace the PID lamp.

 Replace the plug-in VOC sensor.

Stability Fault

Probable Causes: Failed Sensor, empty or close to empty Cal Gas Cylinder, or problems with cal gas and
delivery.

 Check validity of span gas using pull tube or other means (check MFG date on cal gas cylinder).

 Use proper cal gas regulators and tubing for highly corrosive gasses.

 If using Splashguard with Integral Cal Port, must use Calibration Wind Guard or air movement can

compromise span gas delivery.

 Check for obstructions affecting cal gas hitting sensor face (including being wet, blocked, or

corroded).

 Verify adequate Sensor Life.

 Clean or replace the PID lamp.

 Replace the plug-in VOC sensor.

Clearing Fault

Probable Causes: Failed Sensor, Cal Gas not removed at appropriate time, problems with cal gas and delivery,
or background of Target Gas.

 Must recover to < 5% of range in < 5 min after AutoSpan is complete.

 Use bottled air (zero air or N

2

) if there is a known continuous background level.

 Check validity of span gas using pull tube or other means (check MFG date on cal gas cylinder).

 Use proper cal gas regulators and tubing for highly corrosive gasses.

 Check for obstructions affecting cal gas hitting sensor face (including being wet, blocked, or

corroded).

 Verify adequate Sensor Life.

 Clean or replace the PID lamp.

 Replace the plug-in VOC sensor.

Poor Calibration Repeatability

Probable Causes: Failed Sensor, use of wrong Cal Gas or problems with cal gas and delivery, or Interference
Gasses.

 Check for adequate Sensor Life.

 Check validity of span gas using pull tube or other means (check MFG date on cal gas cylinder).

 Use proper cal gas regulators and tubing for highly corrosive gasses.

 Check for obstructions affecting cal gas hitting sensor face (including being wet, blocked, or

corroded).

 Verify adequate Sensor Life.

 Clean or replace the PID lamp.

 Replace the plug-in VOC sensor.

Unstable Output/ Sudden spiking

Possible Causes: Unstable power supply, inadequate grounding, or inadequate RFI protection.

 Verify Power source is stable.