Model PI-700

PI-700 Instruction Manual

Rev. 2.5

Page 23 of 54

The menu item appears as: “Set Zero Offset

From the “Set Zero Offset” text scroll, hold the programming magnet over PGM1 or PGM2 until the “◄”
prompt appears and then hold continue to hold the magnet in place for an additional 3-4 seconds (until the
display starts to scroll “Set Zero Offset”). The display will then switch to “ X.X“ (where X.X is the current
offset). Swipe the magnet momentarily over PGM2 to increase or PGM1 to decrease the number until the
desired zero offset is displayed. Hold the magnet over PGM1 or PGM2 for 3-4 seconds to accept the new
value. The display will scroll “ID Saved”, and revert to “Set Zero Offset” text scroll.

Move to another menu item by executing a momentary hold, or, return to Normal Operation via automatic
timeout of about 15 seconds (the display will scroll “Set Zero Offset” 5 times and then return to Normal
Operation).

If performed correctly the sensor should read 0.0 after returning to normal operation.

3.5.7

Signal Output Check

Signal Output Check provides a simulated 4-20mA output and RS-485 Modbus™ output. This simulation
allows the user to conveniently perform a functional system check of their entire safety system. This signal
output simulation also aids the user in performing troubleshooting of signal wiring problems.

The menu item appears as: “Signal Output Check”.

From the “Signal Output Check” text scroll, hold the magnet over PGM1 or PGM2 until the “◄” prompt
appears and then hold continuously for an additional 10 seconds.

Once initiated, the display will scroll

Simulation Active” until the function is stopped.

During simulation mode, the 4-20mA value will be

increased from 4.0mA to 20.0mA (in 1% of range increments at about a 1 second update rate) and then
decreased from 20.0mA to 4.0mA. The same simulation sequence is applied to the Modbus™ output gas
reading.

NOTE: Signal Output Check stays active indefinitely until the user stops the function. There
is no automatic timeout for this feature.

To end simulation mode, hold magnet over PGM1 or PGM2 for 3 seconds. The display will either move to the
prior menu item or move to the next menu item respectively.

Move to another menu item by executing a momentary hold, or, return to Normal Operation via automatic
timeout of about 15 seconds.

3.5.8

Restore Factory Defaults

Restore Factory Defaults is used to clear current user configuration and calibration data from memory and
revert to factory default values. This may be required if the settings have been configured improperly and a
known reference point needs to be re-established to correct the problem.

This menu item appears as: “Restore Defaults”.

NOTE: Restoring factory defaults should only be used when absolutely necessary.

All

previously existing configuration inputs will have to be re-entered if this function is executed.
A full 10-second magnet hold on PGM 2 is required to execute this function.